Human S100B PicoKine ELISA Kit

Brand:
Boster Bio
Storage:

For quantitative detection of human S100B in cell culture supernates, serum and plasma(heparin, EDTA).   

 

Available on backorder

SKU: EK1163 Category:

Description

Boster’s human human S100B ELISA Kit was based on standard sandwich enzyme-linked immune-sorbent assay technology. A monoclonal antibody from mouse specific for S100B has been precoated onto 96-well plates. Standards(Expression system for standard: E.coli, Immunogen sequence: S2-E92) and test samples are added to the wells, a biotinylated detection polyclonal antibody from goat specific for S100B is added subsequently and then followed by washing with PBS or TBS buffer. Avidin-Biotin-Peroxidase Complex was added and unbound conjugates were washed away with PBS or TBS buffer. HRP substrate TMB was used to visualize HRP enzymatic reaction. TMB was catalyzed by HRP to produce a blue color product that changed into yellow after adding acidic stop solution. The density of yellow is proportional to the human Fibronectin amount of sample captured in plate.


Synonyms: Protein S100-B; S-100 protein beta chain; S-100 protein subunit beta; S100 calcium-binding protein B; S100B

Species: Human S100B

Assay: 156pg/ml-10000pg/ml

Sensitivy: <50pg/ml Specifity: Natural and recombinant human S100B

Cross Reactivity: There is no detectable cross-reactivity with other relevant proteins.

Antibodies: A monoclonal antibody from mouse|a biotinylated detection polyclonal antibody from goat

Anticoagulant: heparin or EDTA

Immunogen: E.coli, S2-E92

Gene name: Protein S100-B / S100 calcium binding protein B

Protein name: Protein S100-B

Storage: Store at 4˚C for 6 months, at -20˚C for 12 months. Avoid multiple freeze-thaw cycles(Shipped with wet ice.)


|Intra-Assay Precision|Inter-Assay Precision
Sample|1|2|3|1|2|3
n|16|16|16|24|24|24
Mean(pg/ml)|228|1186|3717|230|1200|4500
Standard deviation|9.58|52.2|219.3|14.03|86.4|337.5
CV(%)|4.2%|4.4%|5.9%|6.1%|7.2%|7.5%


Reproducibility

To assay reproducibility, three samples with differing target protein concentrations were assayed using four different lots.

Lots Lot1 (pg/ml) Lot2 (pg/ml) Lot3 (pg/ml) Lot4 (pg/ml) Mean (pg/ml) Standard Deviation CV (%)
Sample 1 228 232 224 231 228 3.11 1.3%
Sample 2 1186 1319 1112 1217 1208 74.33 6.1%
Sample 3 3717 3946 3642 4132 3859 193.22 5.0%

*number of samples for each test n=16.


Catalog number|Description|Quantity
EK1163-CAP|96-well plate precoated with anti- human S100B antibody|1
EK1163-ST|Lyophilized recombinant human S100B standard|10ng/tube×2
EK1163-DA|Biotinylated anti- human S100B antibody|130μl(dilution 1:100)
AR1103|Avidin-Biotin-Peroxidase Complex (ABC)|130μl(dilution 1:100)
AR1106-1|Sample diluent buffer|30 ml
AR1106-2|Antibody diluent buffer|12ml
AR1106-3|ABC diluent buffer|12ml
AR1104|TMB color developing agent|10ml
AR1105|TMB stop solution|10ml
PLA-SEA|Adhesive cover|4