Human FOLR1 PicoKine ELISA Kit

Brand:
Boster Bio
Storage:

For quantitative detection of human FOLR1 in cell culture supernates, serum, plasma(heparin, EDTA), saliva, urine and human milk.   

 

Available on backorder

SKU: EK1242 Category:

Description

Boster’s human FOLR1 ELISA Kit was based on standard sandwich enzyme-linked immune-sorbent assay technology. A monoclonal antibody from mouse specific for FOLR1 has been precoated onto 96-well plates. Standards(E.coli, R25-S234) and test samples are added to the wells, a biotinylated detection polyclonal antibody from goat specific for FOLR1 is added subsequently and then followed by washing with PBS or TBS buffer. Avidin-Biotin-Peroxidase Complex was added and unbound conjugates were washed away with PBS or TBS buffer. HRP substrate TMB was used to visualize HRP enzymatic reaction. TMB was catalyzed by HRP to produce a blue color product that changed into yellow after adding acidic stop solution. The density of yellow is proportional to the human FOLR1 amount of sample captured in plate.


Synonyms: Adult folate binding protein|Adult folate-binding protein|FBP|Folate Receptor 1 Adult|Folate receptor 1|Folate Receptor 1 Precursor|Folate receptor adult|Folate receptor alpha|Folate receptor|FOLR|FOLR1|FOLR1_HUMAN|FR alpha|FR-alpha|FRalpha|KB cells FBP|MOV18|Ovarian cancer associated antigen|Ovarian tumor associated antigen|Ovarian tumor associated antigen MOv18|Ovarian tumor-associated antigen MOv18

Species: human FOLR1

Assay: 46.9pg/ml-3000pg/ml

Sensitivy: <10pg/ml Specifity: Natural and recombinant human FOLR1

Cross Reactivity: There is no detectable cross-reactivity with other relevant proteins.

Antibodies: a monoclonal antibody from mouse|a biotinylated polyclonal antibody from goat

Anticoagulant: heparin or EDTA

Immunogen: E.coli, R25-S234

Gene name: Folate receptor alpha / folate receptor 1 (adult)

Protein name: Folate receptor alpha

Storage: Store at 4˚C for 6 months, at -20˚C for 12 months. Avoid multiple freeze-thaw cycles(Shipped with wet ice.)


|Intra-Assay Precision|Inter-Assay Precision
Sample|1|2|3|1|2|3
n|16|16|16|24|24|24
Mean(pg/ml)|103|333|1306|108|312|1388
Standard deviation|7.41|22.31|88.8|8.53|24.96|119.36
CV(%)|7.2%|6.7%|6.8%|7.9%|8%|8.6%


Reproducibility

To assay reproducibility, three samples with differing target protein concentrations were assayed using four different lots.

Lots Lot1 (pg/ml) Lot2 (pg/ml) Lot3 (pg/ml) Lot4 (pg/ml) Mean (pg/ml) Standard Deviation CV (%)
Sample 1 103 108 100 99 102 3.5 3.4%
Sample 2 333 359 355 339 346 10.8 3.1%
Sample 3 1306 1386 1411 1506 1402 71.35 5%

*number of samples for each test n=16.


Catalog number|Description|Quantity
EK1242-CAP|96-well plate precoated with anti- human FOLR1 antibody|1
EK1242-ST|Lyophilized recombinant human FOLR1 standard|10ng/tube×2
EK1242-DA|Biotinylated anti- human FOLR1 antibody|130μl(dilution 1:100)
AR1103|Avidin-Biotin-Peroxidase Complex (ABC)|130μl(dilution 1:100)
AR1106-1|Sample diluent buffer|30ml
AR1106-2|Antibody diluent buffer|12ml
AR1106-3|ABC diluent buffer|12ml
AR1104|TMB color developing agent|10ml
AR1105|TMB stop solution|10ml
PLA-SEA|Adhesive cover|4