Description
Boster’s human TIMP-1 ELISA Kit was based on standard sandwich enzyme-linked immune-sorbent assay technology. A monoclonal antibody from mouse specific for TIMP-1 has been precoated onto 96-well plates. Standards(NSO, C24-A207) and test samples are added to the wells, a biotinylated detection polyclonal antibody from goat specific for TIMP-1 is added subsequently and then followed by washing with PBS or TBS buffer. Avidin-Biotin-Peroxidase Complex was added and unbound conjugates were washed away with PBS or TBS buffer. HRP substrate TMB was used to visualize HRP enzymatic reaction. TMB was catalyzed by HRP to produce a blue color product that changed into yellow after adding acidic stop solution. The density of yellow is proportional to the human TIMP-1 amount of sample captured in plate.
Synonyms: Metalloproteinase inhibitor 1;Erythroid-potentiating activity;EPA;Fibroblast collagenase inhibitor;Collagenase inhibitor;Tissue inhibitor of metalloproteinases 1;TIMP-1;TIMP1;CLGI, TIMP;
Species: Human TIMP-1
Assay: 156pg/ml-10,000pg/ml
Sensitivy: <5pg/ml Specifity: Natural and recombinant human TIMP-1
Cross Reactivity: There is no detectable cross-reactivity with other relevant proteins.
Antibodies: monoclonal antibody from mouse|polyclonal antibody from goat
Anticoagulant: heparin or EDTA
Immunogen: Expression system for standard: NSO; Immunogen sequence: C24-A207
Gene name: Metalloproteinase inhibitor 1 / Metalloproteinase inhibitor 1
Protein name: Metalloproteinase inhibitor 1
Storage: Store at 4°C for 6 months, at -20°C for 12 months. Avoid multiple freeze-thaw cycles(Shipped with wet ice.)
Intra-Assay Precision | Inter-Assay Precision | |||||
---|---|---|---|---|---|---|
Sample | 1 | 2 | 3 | 1 | 2 | 3 |
n | 16 | 16 | 16 | 24 | 24 | 24 |
Mean(pg/ml) | 277 | 1942 | 4295 | 285 | 2099 | 4333 |
Standard deviation | 17.4 | 122.34 | 292.06 | 21.66 | 146.93 | 363.97 |
CV(%) | 6.8% | 7% | 6.8% | 7.6% | 7% | 8.4% |
Reproducibility
To assay reproducibility, three samples with differing target protein concentrations were assayed using four different lots.
Lots | Lot1 (pg/ml) | Lot2 (pg/ml) | Lot3 (pg/ml) | Lot4 (pg/ml) | Mean (pg/ml) | Standard Deviation | CV (%) |
---|---|---|---|---|---|---|---|
Sample 1 | 277 | 258 | 271 | 245 | 262 | 12.33 | 4.7% |
Sample 2 | 1942 | 2029 | 1819 | 1993 | 1945 | 79.43 | 4% |
Sample 3 | 4295 | 4957 | 4587 | 4338 | 4544 | 263.08 | 5.7% |
*number of samples for each test n=16.
10ng/tube