Description
Boster’s human sL-Selectin ELISA Kit was based on standard sandwich enzyme-linked immune-sorbent assay technology. A monoclonal antibody from mouse specific for sL-Selectin has been precoated onto 96-well plates. Standards(NSO, W39-N332) and test samples are added to the wells, a biotinylated detection polyclonal antibody from goat specific for sL-Selectin is added subsequently and then followed by washing with PBS or TBS buffer. Avidin-Biotin-Peroxidase Complex was added and unbound conjugates were washed away with PBS or TBS buffer. HRP substrate TMB was used to visualize HRP enzymatic reaction. TMB was catalyzed by HRP to produce a blue color product that changed into yellow after adding acidic stop solution. The density of yellow is proportional to the human sL-Selectin amount of sample captured in plate.
Synonyms: A.11|AI528707|CD62 antigen ligand|CD62 antigen-like family member L|CD62L antigen|gp90 MEL|hLHRc|IgA nephropathy, susceptibility to, included|L Selectin|L selectin precursor|LAM 1|LAM1|LECAM1|Leu 8|Leu8|Leukocyte adhesion molecule 1|Leukocyte endothelial cell adhesion molecule 1|Leukocyte surface antigen Leu 8|Leukocyte surface antigen Leu8|Leukocyte-endothelial cell adhesion molecule 1|LNHR|LSEL|Ly-22|Ly-m22|Lyam 1|LYAM1|Lymph node homing receptor|Lymphocyte adhesion molecule 1|Lymphocyte antigen 22|Lymphocyte surface MEL-14 antigen|PLNHR|Selectin L|Selectin L (lymphocyte adhesion molecule 1)|Selectin, lymphocyte|SELL|TQ1
Species: Human sL-Selectin
Assay: 62.5pg/ml-4000pg/ml
Sensitivy: <5pg/ml Specifity: Natural and recombinant human sL-Selectin
Cross Reactivity: There is no detectable cross-reactivity with other relevant proteins.
Antibodies: a monoclonal antibody from mouse|a biotinylated polyclonal antibody from goat
Anticoagulant: heparin, EDTA or citrate
Immunogen: NSO, W39-N332
Gene name: L-selectin / selectin L
Protein name: L-selectin
Storage: Store at 4˚C for 6 months, at -20˚C for 12 months. Avoid multiple freeze-thaw cycles(Shipped with wet ice.)
|Intra-Assay Precision|Inter-Assay Precision
Sample|1|2|3|1|2|3
n|16|16|16|24|24|24
Mean(pg/ml)|195|543|1239|187|595|1352
Standard deviation|15.21|33.66|60.71|16.64|37.48|67.6
CV(%)|7.8%|6.2%|4.9%|8.9%|6.3%|5%
Reproducibility
To assay reproducibility, three samples with differing target protein concentrations were assayed using four different lots.
Lots | Lot1 (pg/ml) | Lot2 (pg/ml) | Lot3 (pg/ml) | Lot4 (pg/ml) | Mean (pg/ml) | Standard Deviation | CV (%) |
---|---|---|---|---|---|---|---|
Sample 1 | 195 | 209 | 190 | 190 | 196 | 7.77 | 3.9% |
Sample 2 | 543 | 572 | 564 | 574 | 563 | 12.27 | 2.1% |
Sample 3 | 1239 | 1413 | 1184 | 1268 | 1276 | 84.65 | 6.6% |
*number of samples for each test n=16.
Catalog number|Description|Quantity
EK0503-CAP|96-well plate precoated with anti- human sL-Selectin antibody|1
EK0503-ST|Lyophilized recombinant human sL-Selectin standard|10ng/tubex2
EK0503-DA|Biotinylated anti- human sL-Selectin antibody|130ul(dilution 1:100)
AR1103|Avidin-Biotin-Peroxidase Complex(ABC)|130ul(dilution 1:100)
AR1106-1|Sample diluent buffer|30ml
AR1106-2|Antibody diluent buffer|12ml
AR1106-3|ABC diluent buffer|12ml
AR1104|TMB color developing agent|10ml
AR1105|TMB stop solution|10ml
PLA-SEA|Adhesive cover|4