Human CD105 PicoKine™ Fast ELISA Kit

Brand:
Boster Bio
Storage:

The Fast version of Picokine ELISA kits, assay takes less than 1.5 hours. Detect Human Endoglin/ENG with <15pg/ml sensitivity. Format: 96-well plate with removable strips. Compatible samples: cell culture supernates, cell lysates, serum and plasma (heparin, EDTA). This is a TMB colorimetric sandwich ELISA kit with short assay time and fast experiment set up. Endoglin/ENG tissue specificity: Endoglin is restricted to endothelial cells in all tissues except bone marrow.   

 

Available on backorder

SKU: FEK0644 Category:

Description

Boster’s human CD105 ELISA Kit was based on standard sandwich enzyme-linked immune-sorbent assay technology. A monoclonal antibody from mouse specific for CD105 has been precoated onto 96-well plates. Standards(CHO, E26-G586) and test samples are added to the wells, a biotinylated detection polyclonal antibody from goat specific for CD105 is added subsequently and then followed by washing with PBS or TBS buffer. Avidin-Biotin-Peroxidase Complex was added and unbound conjugates were washed away with PBS or TBS buffer. HRP substrate TMB was used to visualize HRP enzymatic reaction. TMB was catalyzed by HRP to produce a blue color product that changed into yellow after adding acidic stop solution. The density of yellow is proportional to the human CD105 amount of sample captured in plate.


Synonyms: Endoglin;CD105;ENG;END;

Species: Human CD105

Assay: 156pg/ml-10,000pg/ml

Sensitivy: <15pg/ml Specifity: Natural and recombinant human CD105

Cross Reactivity: There is no detectable cross-reactivity with other relevant proteins.

Antibodies: monoclonal antibody from mouse|polyclonal antibody from goat

Anticoagulant: heparin or EDTA

Immunogen: Expression system for standard: CHO; Immunogen sequence: E26-G586

Gene name: Endoglin / Endoglin

Protein name: Endoglin

Storage: Store at 4°C for 6 months, at -20°C for 12 months. Avoid multiple freeze-thaw cycles(Shipped with wet ice.)


Intra-Assay Precision Inter-Assay Precision
Sample 1 2 3 1 2 3
n 16 16 16 24 24 24
Mean(pg/ml) 312 1384 4821 316 1275 4955
Standard deviation 16.22 74.73 318.18 17.38 72.67 386.49
CV(%) 5.2% 5.4% 6.6% 5.5% 5.7% 7.8%

Reproducibility

To assay reproducibility, three samples with differing target protein concentrations were assayed using four different lots.

Lots Lot1 (pg/ml) Lot2 (pg/ml) Lot3 (pg/ml) Lot4 (pg/ml) Mean (pg/ml) Standard Deviation CV (%)
Sample 1 312 285 280 306 295 13.53 4.5%
Sample 2 1384 1301 1412 1238 1333 68.71 5.1%
Sample 3 4821 5271 4779 5474 5086 295.48 5.8%

*number of samples for each test n=16.


10ng/tube