Rat CD44 PicoKine™ Fast ELISA Kit

Brand:
Boster Bio
Storage:

The Fast version of Picokine ELISA kits, assay takes less than 1.5 hours. Detect Rat Cd44 with <10pg/ml sensitivity. Format: 96-well plate with removable strips. Compatible samples: cell culture supernates, cell lysates, serum and plasma (heparin, EDTA). This is a TMB colorimetric sandwich ELISA kit with short assay time and fast experiment set up. Cd44 tissue specificity:   

 

Available on backorder

SKU: FEK1417 Category:

Description

Boster’s rat CD44 ELISA Kit was based on standard sandwich enzyme-linked immune-sorbent assay technology. A monoclonal antibody from mouse specific for CD44 has been precoated onto 96-well plates. Standards(NSO, Q22 – E271) and test samples are added to the wells, a biotinylated detection polyclonal antibody from goat specific for CD44 is added subsequently and then followed by washing with PBS or TBS buffer. Avidin-Biotin-Peroxidase Complex was added and unbound conjugates were washed away with PBS or TBS buffer. HRP substrate TMB was used to visualize HRP enzymatic reaction. TMB was catalyzed by HRP to produce a blue color product that changed into yellow after adding acidic stop solution. The density of yellow is proportional to the rat CD44 amount of sample captured in plate.


Synonyms: CD44 antigen;Extracellular matrix receptor III;ECMR-III;GP90 lymphocyte homing/adhesion receptor;HUTCH-I;Hermes antigen;Hyaluronate receptor;Phagocytic glycoprotein 1;PGP-1;Phagocytic glycoprotein I;PGP-I;CD44;Cd44;

Species: rat CD44

Assay: 312pg/ml-20000pg/ml

Sensitivy: <10pg/ml Specifity: Natural and recombinant rat CD44

Cross Reactivity: There is no detectable cross-reactivity with other relevant proteins.

Antibodies: monoclonal antibody from mouse|polyclonal antibody from goat

Anticoagulant: heparin or EDTA

Immunogen: Expression system for standard: NSO; Immunogen sequence: Q22-E271

Gene name: CD44 antigen / CD44 antigen

Protein name: CD44 antigen

Storage: Store at 4°C for 6 months, at -20°C for 12 months. Avoid multiple freeze-thaw cycles(Shipped with wet ice.)


Intra-Assay Precision Inter-Assay Precision
Sample 1 2 3 1 2 3
n 16 16 16 24 24 24
Mean(pg/ml) 481 1665 7855 452 1537 7710
Standard deviation 28.37 129.87 439.88 32.54 136.79 516.57
CV(%) 5.6% 7.8% 5.6% 7.2% 8.9% 6.7%

Reproducibility

To assay reproducibility, three samples with differing target protein concentrations were assayed using four different lots.

Lots Lot1 (pg/ml) Lot2 (pg/ml) Lot3 (pg/ml) Lot4 (pg/ml) Mean (pg/ml) Standard Deviation CV (%)
Sample 1 481 494 514 475 491 14.94 3%
Sample 2 1665 1566 1588 1645 1616 40.39 2.4%
Sample 3 7855 9238 8886 9256 8808 570.06 6.4%

*number of samples for each test n=16.


20ng/tube