Human CD56/NCAM-1 PicoKine ELISA Kit

Brand:
Boster Bio
Storage:

For quantitative detection of human CD56 in cell culture supernates, cell lysates, serum and plasma (heparin, EDTA).   

 

Available on backorder

SKU: EK0706 Category:

Description

Boster’s human CD56 ELISA Kit was based on standard sandwich enzyme-linked immune-sorbent assay technology. A monoclonal antibody from mouse specific for CD56 has been precoated onto 96-well plates. Standards(NSO, L20-G718) and test samples are added to the wells, a biotinylated detection polyclonal antibody from goat specific for CD56 is added subsequently and then followed by washing with PBS or TBS buffer. Avidin-Biotin-Peroxidase Complex was added and unbound conjugates were washed away with PBS or TBS buffer. HRP substrate TMB was used to visualize HRP enzymatic reaction. TMB was catalyzed by HRP to produce a blue color product that changed into yellow after adding acidic stop solution. The density of yellow is proportional to the human CD56 amount of sample captured in plate.


Synonyms: CD56 | NCAM 1 | NCAM C | NCAM1 | NCAM-1 | N-CAM-1 | P13591

Species: Human CD56

Assay: 156pg/ml-10000pg/ml

Sensitivy: <10pg/ml Specifity: Natural and recombinant human CD56

Cross Reactivity: There is no detectable cross-reactivity with other relevant proteins.

Antibodies: a monoclonal antibody from mouse|a biotinylated polyclonal antibody from goat

Anticoagulant: heparin or EDTA

Immunogen: NSO, L20-G718

Gene name: Neural cell adhesion molecule 1 / neural cell adhesion molecule 1

Protein name: Neural cell adhesion molecule 1

Storage: Store at 4˚C for 6 months, at -20˚C for 12 months. Avoid multiple freeze-thaw cycles(Shipped with wet ice.)


|Intra-Assay Precision|Inter-Assay Precision
Sample|1|2|3|1|2|3
n|16|16|16|24|24|24
Mean(pg/ml)|319|1085|5570|313|1102|5558
Standard deviation|16.26|44.48|239.51|17.21|50.69|316.8
CV(%)|5.1%|4.1%|4.3%|5.5%|4.6%|5.7%


Reproducibility

To assay reproducibility, three samples with differing target protein concentrations were assayed using four different lots.

Lots Lot1 (pg/ml) Lot2 (pg/ml) Lot3 (pg/ml) Lot4 (pg/ml) Mean (pg/ml) Standard Deviation CV (%)
Sample 1 319 341 313 295 317 16.43 5.1%
Sample 2 1085 1189 1030 1018 1080 67.54 6.2%
Sample 3 5570 5244 5973 5762 5637 268.07 4.7%

*number of samples for each test n=16.


Catalog number|Description|Quantity
EK0706-CAP|96-well plate precoated with anti- human CD56 antibody|1
EK0706-ST|Lyophilized recombinant human CD56 standard|10ng/tube×2
EK0706-DA|Biotinylated anti- human CD56 antibody|130μl(dilution 1:100)
AR1103|Avidin-Biotin-Peroxidase Complex (ABC)|130μl(dilution 1:100)
AR1106-1|Sample diluent buffer|30 ml
AR1106-2|Antibody diluent buffer|12ml
AR1106-3|ABC diluent buffer|12ml
AR1104|TMB color developing agent|10ml
AR1105|TMB stop solution|10ml
PLA-SEA|Adhesive cover|4