Description
Boster’s human BMP-2 ELISA Kit was based on standard sandwich enzyme-linked immune-sorbent assay technology. A monoclonal antibody from mouse specific for BMP-2 has been precoated onto 96-well plates. Standards(CHO, Q283-R396) and test samples are added to the wells, a biotinylated detection monoclonal antibody from mouse specific for BMP-2 is added subsequently and then followed by washing with PBS or TBS buffer. Avidin-Biotin-Peroxidase Complex was added and unbound conjugates were washed away with PBS or TBS buffer. HRP substrate TMB was used to visualize HRP enzymatic reaction. TMB was catalyzed by HRP to produce a blue color product that changed into yellow after adding acidic stop solution. The density of yellow is proportional to the human BMP-2 amount of sample captured in plate.
Synonyms: Bone morphogenetic protein 2 ;BMP2 ;
Species: Human BMP-2
Assay: 31.2pg/ml-2000pg/ml
Sensitivy: <2pg/ml Specifity: Natural and recombinant human BMP-2
Cross Reactivity: There is no detectable cross-reactivity with other relevant proteins.
Antibodies: monoclonal antibody from mouse|monoclonal antibody from mouse
Anticoagulant:
Immunogen: Expression system for standard: CHO; Immunogen sequence: Q283-R396
Gene name: Bone morphogenetic protein 2 / bone morphogenetic protein 2
Protein name: Bone morphogenetic protein 2
Storage: Store at 4°C for 6 months, at -20°C for 12 months. Avoid multiple freeze-thaw cycles(Shipped with wet ice.)
Intra-Assay Precision | Inter-Assay Precision | |||||
---|---|---|---|---|---|---|
Sample | 1 | 2 | 3 | 1 | 2 | 3_x000D_ |
n | 16 | 16 | 16 | 24 | 24 | 24_x000D_ |
Mean(pg/ml) | 55 | 250 | 1005 | 59 | 255 | 1061_x000D_ |
Standard deviation | 4.29 | 13.25 | 70.35 | 4.66 | 15.55 | 83.81_x000D_ |
CV(%) | 7.8% | 5.3% | 7% | 7.9% | 6.1% | 7.9% |
Reproducibility
To assay reproducibility, three samples with differing target protein concentrations were assayed using four different lots.
Lots | Lot1 (pg/ml) | Lot2 (pg/ml) | Lot3 (pg/ml) | Lot4 (pg/ml) | Mean (pg/ml) | Standard Deviation | CV (%) |
---|---|---|---|---|---|---|---|
Sample 1 | 55 | 51 | 58 | 51 | 53 | 2.94 | 5.5% |
Sample 2 | 250 | 245 | 240 | 275 | 252 | 13.46 | 5.3% |
Sample 3 | 1005 | 885 | 885 | 952 | 931 | 50.36 | 5.4% |
*number of samples for each test n=16.
10ng/tube