Human ACE PicoKine ELISA Kit

Brand:
Boster Bio
Storage:

For quantitative detection of human ACE in cell culture supernates, serum, plasma(heparin) and saliva.   

 

Available on backorder

SKU: EK0557 Category:

Description

Boster’s human ACE ELISA Kit was based on standard sandwich enzyme-linked immune-sorbent assay technology. A monoclonal antibody from mouse specific for ACE has been precoated onto 96-well plates. Standards(NSO, L30-L1261) and test samples are added to the wells, a biotinylated detection polyclonal antibody from goat specific for ACE is added subsequently and then followed by washing with PBS or TBS buffer. Avidin-Biotin-Peroxidase Complex was added and unbound conjugates were washed away with PBS or TBS buffer. HRP substrate TMB was used to visualize HRP enzymatic reaction. TMB was catalyzed by HRP to produce a blue color product that changed into yellow after adding acidic stop solution. The density of yellow is proportional to the human ACE amount of sample captured in plate.


Synonyms: ACE 1 | ACE T | Angiotensin converting enzyme | Angiotensin Converting Enzyme 1 | Angiotensinconverting enzyme | Angiotensin-converting enzyme | CD143 antigen | DCP 1 | DCP, DCP1 | Dipeptidyl carboxypeptidase 1 | Dipeptidyl carboxypeptidase I | Kininase II | MVCD3 | P12821 | Peptidyl dipeptidase A

Species: Human ACE

Assay: 0.78ng/ml-50ng/ml

Sensitivy: <5pg/ml Specifity: Natural and recombinant human ACE

Cross Reactivity: There is no detectable cross-reactivity with other relevant proteins.

Antibodies: a monoclonal antibody from mouse|a biotinylated polyclonal antibody from goat

Anticoagulant: heparin

Immunogen: NSO, L30-L1261

Gene name: Angiotensin-converting enzyme / angiotensin I converting enzyme

Protein name: Angiotensin-converting enzyme

Storage: Store at 4˚C for 6 months, at -20˚C for 12 months. Avoid multiple freeze-thaw cycles(Shipped with wet ice.)


|Intra-Assay Precision|Inter-Assay Precision
Sample|1|2|3|1|2|3
n|16|16|16|24|24|24
Mean(pg/ml)|1278|6839|24157|1358|6350|25528
Standard deviation|62.62|465.05|1087.06|88.27|482.6|1225.34
CV(%)|4.9%|6.8%|4.5%|6.5%|7.6%|4.8%


Reproducibility

To assay reproducibility, three samples with differing target protein concentrations were assayed using four different lots.

Lots Lot1 (pg/ml) Lot2 (pg/ml) Lot3 (pg/ml) Lot4 (pg/ml) Mean (pg/ml) Standard Deviation CV (%)
Sample 1 1278 1310 1292 1425 1326 58.13 4.3%
Sample 2 6839 6994 7039 6669 6885 145.23 2.1%
Sample 3 24157 24884 22207 21600 23212 1350.66 5.8%

*number of samples for each test n=16.


Catalog number|Description|Quantity
EK0557-CAP|96-well plate precoated with anti- human ACE antibody|1
EK0557-ST|Lyophilized recombinant human ACE standard|50ng/tubex2
EK0557-DA|Biotinylated anti- human ACE antibody|130ul(dilution 1:100)
AR1103|Avidin-Biotin-Peroxidase Complex(ABC)|130ul(dilution 1:100)
AR1106-1|Sample diluent buffer|30ml
AR1106-2|Antibody diluent buffer|12ml
AR1106-3|ABC diluent buffer|12ml
AR1104|TMB color developing agent|10ml
AR1105|TMB stop solution|10ml
PLA-SEA|Adhesive cover|4