Human CDCP1 PicoKine ELISA Kit

Brand:
Boster Bio
Storage:

For Quantitative Detection of human CDCP1 in cell culture supernates, serum and plasma (heparin, EDTA, citrate).   

 

Available on backorder

SKU: EK1668 Category:

Description

Boster’s human CDCP1 ELISA Kit was based on standard sandwich enzyme-linked immune-sorbent assay technology. A monoclonal antibody from mouse specific for CDCP1 has been precoated onto 96-well plates. Standards(Expression system for standard: NSO, Immunogen sequence: F30-T667) and test samples are added to the wells, a biotinylated detection polyclonal antibody from goat specific for CDCP1 is added subsequently and then followed by washing with PBS or TBS buffer. Avidin-Biotin-Peroxidase Complex was added and unbound conjugates were washed away with PBS or TBS buffer. HRP substrate TMB was used to visualize HRP enzymatic reaction. TMB was catalyzed by HRP to produce a blue color product that changed into yellow after adding acidic stop solution. The density of yellow is proportional to the human CDCP1 amount of sample captured in plate.


Synonyms: CUB domain-containing protein 1; Membrane glycoprotein gp140; Subtractive immunization M plus HEp3-associated 135 kDa protein; SIMA135; Transmembrane and associated with src kinases; CD318; CDCP1; TRASK; UNQ2486/PRO5773

Species: Human CDCP1

Assay: 31.2pg/ml-2000pg/ml

Sensitivy: <2pg/ml Specifity: Natural and recombinant human CDCP1

Cross Reactivity: There is no detectable cross-reactivity with other relevant proteins.

Antibodies: a monoclonal antibody from mouse|a biotinylated polyclonal antibody from goat

Anticoagulant: heparin, EDTA or citrate

Immunogen: NSO,F30-T667

Gene name: CUB domain-containing protein 1 / CUB domain containing protein 1

Protein name: CUB domain-containing protein 1

Storage: Store at 4˚C for 6 months, at -20˚C for 12 months. Avoid multiple freeze-thaw cycles(Shipped with wet ice.)


|Intra-Assay Precision|Inter-Assay Precision
Sample|1|2|3|1|2|3
n|16|16|16|24|24|24
Mean(pg/ml)|54|307|865|54|291|945
Standard deviation|4.26|12.89|62.28|4.59|14.84|68.98
CV(%)|7.9%|4.2%|7.2%|8.5%|5.1%|7.3%


Reproducibility

To assay reproducibility, three samples with differing target protein concentrations were assayed using four different lots.

Lots Lot1 (pg/ml) Lot2 (pg/ml) Lot3 (pg/ml) Lot4 (pg/ml) Mean (pg/ml) Standard Deviation CV (%)
Sample 1 54 53 62 62 57 4.26 7.4%
Sample 2 307 305 306 280 299 11.28 3.7%
Sample 3 865 848 944 941 899 43.43 4.8%

*number of samples for each test n=16.


Catalog number|Description|Quantity
EK1668-CAP|96-well plate precoated with anti- human CDCP1 antibody|1
EK1668-ST|Lyophilized recombinant human CDCP1 standard|2ng/tube×2
EK1668-DA|Biotinylated anti- human CDCP1 antibody|130μl(dilution 1:100)
AR1103|Avidin-Biotin-Peroxidase Complex (ABC)|130μl(dilution 1:100)
AR1106-1|Sample diluent buffer|30 ml
AR1106-2|Antibody diluent buffer|12ml
AR1106-3|ABC diluent buffer|12ml
AR1104|TMB color developing agent|10ml
AR1105|TMB stop solution|10ml
PLA-SEA|Adhesive cover|4